Automatic regulatory control in type 1 diabetes without carbohydrate counting

Colmegna, Patricio
Garelli, Fabricio
De Battista, Hernán
Sánchez-Peña, Ricardo
Type
Artículo de publicación periódica
Citations
Altmetric:
Abstract
"A new approach to automatically regulate the glucose level in type 1 diabetes is presented in this work. This is the so-called Automatic Regulation of Glucose (ARG) algorithm, which is based on a switched Linear Quadratic Gaussian (LQG) inner controller, combined with an outer sliding mode safety loop with Insulin on Board (IOB) constraints. In silico and in vivo results without feedforward insulin boluses delivered at meal times indicate that safe blood glucose control can be achieved by the proposed controller. This controller is simple to migrate to well-known hardware platforms, and intuitive to tune using a priori clinical information."
Description
Date
2018-05
Journal Title
Journal ISSN
Volume Title
Publisher
Research Projects
Organizational Units
Journal Issue
Keywords
DIABETES, CONTROL AUTOMATICO, ORGANOS ARTIFICIALES, CONTROL EN MODO DESLIZANTE, GLUCOSA, PANCREAS
Citation